Common Functions
© VeEX Inc. All Rights Reserved.
296
l
Zero
: Recalibrates the OPM to treat current conditions as zero. Used when measurement conditions
change significantly or to re-calibrate the OPM. Make sure the cover is shut on the OPM test port
before tapping
Zero
. When in doubt, recalibrate prior to making any measurements, e.g, when test-
ing in cold outdoor temperatures and then moving testing into a heated building.
l
OPM Server On:
Remote control feature not currently supported.
If your device is equipped with a built-in clock, the saved measurement result will have a
time stamp.
To permanently save the table of results, you must save into a test file.
l
Mode
: Automatically designates the type of OPM: PM1 (-70 to +10dBm), PM2 (-50 to +25dBm), or
PM3 (-65 to +15dBm).
No user selection available.
Results
Results can be saved to the test set, exported to USB, or uploaded to Fiberizer Cloud
Results Management Menu
To view and save (enter) test results into the Results table:
1. Select the OPM tab. Verify the insertion loss (IL) values are acceptable, then tap
Acquire
to load val-
ues into the
Table
tab. The
Table
tab title shows how many fiber records have been saved in the
Содержание MTX150 Series
Страница 1: ...MTX150x User Manual Multi Service Installation Maintenance Test Set P N D07 00 148P Rev B00 ...
Страница 11: ...Customer Care 303 Table of Contents VeEX Inc All Rights Reserved 11 ...
Страница 171: ...171 VeEX Inc All Rights Reserved Ethernet V FTP Setup Download V FTP Download Results SyncE ...
Страница 246: ...Common Functions VeEX Inc All Rights Reserved 246 Bluetooth Devices Bluetooth Scan ...
Страница 248: ...Common Functions VeEX Inc All Rights Reserved 248 Bluetooth Devices Bluetooth Scan ...
Страница 267: ...267 VeEX Inc All Rights Reserved Common Functions ...
Страница 300: ...Software Update Pro cess VeEX Inc All Rights Reserved 300 Software Version is displayed in the About section ...