Ethernet
l
Availability:
Minimum percentage of service availability allowed to still be compliant with the SLA.
The service becomes unavailable if more than 50% of the frames are errored or missing in a one
second interval. Availability is only guaranteed for traffic conforming to the CIR.
l
Total RX Frames:
Total number of frames received
l
Errored Frame Count:
Number of frames with CRC or IP Checksum errors
Measured values that do not meet the service test parameters set in the Bandwidth and Threshold tabs cause
the test to fail.
The
Summary
tab displays the status of each service and test as
Pass
,
Failed
,
Pending
, or
Disabled
.
Perf. Tests - Summary
Signal tab:
The Signal tab (fiber ports only) displays the optical level measured by the SFP transceiver. The
RFC 2544 Signal tab is identical to the Signal tab for the BERT application. Refer to
from the BERT sec-
tion for more information.
Perf. Tests - Summary
Содержание MTX150 Series
Страница 1: ...MTX150x User Manual Multi Service Installation Maintenance Test Set P N D07 00 148P Rev B00 ...
Страница 11: ...Customer Care 303 Table of Contents VeEX Inc All Rights Reserved 11 ...
Страница 171: ...171 VeEX Inc All Rights Reserved Ethernet V FTP Setup Download V FTP Download Results SyncE ...
Страница 246: ...Common Functions VeEX Inc All Rights Reserved 246 Bluetooth Devices Bluetooth Scan ...
Страница 248: ...Common Functions VeEX Inc All Rights Reserved 248 Bluetooth Devices Bluetooth Scan ...
Страница 267: ...267 VeEX Inc All Rights Reserved Common Functions ...
Страница 300: ...Software Update Pro cess VeEX Inc All Rights Reserved 300 Software Version is displayed in the About section ...