Ethernet
© VeEX Inc. All Rights Reserved.
106
Select Simple Test, Step Load Test, or Simple and Step.
l
Simple Test:
Starts the tests at the CIR.
l
Step Load Test:
Starts the test below the CIR and continues in steps until it reaches the CIR.
l
Simple and Step Load Test:
Step Load Test performs only if the Simple Validation test fails.
Tap the
Test Duration
box to input a test duration (test duration must be less than 999 sec).
Tap the table to modify the CIR value percentage for each step.
CIR Test Config
Header Settings
l
Service #:
Select a service to configure
l
Service Name:
Assign a name to the service if desired.
l
Frame Size Type:
Fixed or EMIX (1GE only). A fixed frame size is chosen as default
l
Frame Size:
l
For Fixed Traffic Flow:
Input a fixed frame size within the range of 64-10000 bytes by tap-
ping the value box.
l
For EMIX (1GE only):
The default value is abceg. Tap the zoom (magnifying glass) icon to
define other values. Select the values from the drop down lists on the next screen.
Any EMIX configuration of 5 frames is allowed.
Содержание MTX150 Series
Страница 1: ...MTX150x User Manual Multi Service Installation Maintenance Test Set P N D07 00 148P Rev B00 ...
Страница 11: ...Customer Care 303 Table of Contents VeEX Inc All Rights Reserved 11 ...
Страница 171: ...171 VeEX Inc All Rights Reserved Ethernet V FTP Setup Download V FTP Download Results SyncE ...
Страница 246: ...Common Functions VeEX Inc All Rights Reserved 246 Bluetooth Devices Bluetooth Scan ...
Страница 248: ...Common Functions VeEX Inc All Rights Reserved 248 Bluetooth Devices Bluetooth Scan ...
Страница 267: ...267 VeEX Inc All Rights Reserved Common Functions ...
Страница 300: ...Software Update Pro cess VeEX Inc All Rights Reserved 300 Software Version is displayed in the About section ...