TECHNICAL INFORMATION
57
87-900-121-01 (
B
)
•
Parameter Number:
it points out to the controller
about which parameter the master wants to operate,
and about the kind of operation (read/write). It
consists of:
1
2
3
4
5
0=
parameter
reading
1 =
parameter
writing
always 0
000 to 299 = commands
300 to 699 = status
700 to 999 = parameter
•
Data Field Length:
it points out the next field length
in characters. In this protocol the possible
combinations are:
02 = parameter reading request
06 = parameter, writing
•
Data:
in this controller it can be:
“=?”:
in the case of parameter reading request
"000000"
or
"111111":
in the case of YES/NO
command or parameter writing
a six numeric character string 0 to 9
pointing out the value of a configuration parameter
or an analog measure
a six character string
pointing out an error
•
Checksum:
it is the sum of the string ASCII values
up to the first character of the checksum with a 8 bit
cutting-off and a conversion to a decimal number.
•
CR:
< 0xD>
All the fields are set on the right with the "0" character
used as filling character.
When a string with an address number equal to the
card one (001 in this case) is received, the controller
answers as detailed in the following:
•
in the case of a parameter request with a string
having the same size described before, but the first
character of the "Parameter Number" field set to 1.
Example for a low speed status request:
Master to Controller:
0010000202= ?097< 0xD >
Controller to Master:
0011000206111111016< 0xD >
where " 111111" is for low speed status ON
•
in the case of a parameter writing, with a string
including the parameter value that has been set out.
Example for a low speed status ON writing:
Master to Controller:
0011000206111111016< 0xD
>
Controller to Master:
0011000206111111016< 0xD
>
•
in the case of a string with a length greater than 40
characters, with:
N
A
K
CR
Controller address
•
in the case of a wrong checksum, with:
N
A
K
CR
Controller address
•
in the case of a parameter number not present,
with:
1 0
0 6 N O - D E F
control.
address
param.
address
check-
sum
CR
•
in the case of an out-limit value, with:
1 0
0 6 - R A N G E
control.
address
param.
address
check-
sum
CR
•
in the case of an attempt of writing a logic
parameter with a value not equal to " 000000" or "
111111", or when the parameter is a read only one,
with:.
1 0
0 6 - L O G I C
control.
address
param.
address
check-
sum
CR
•
in the case of a time between two characters
greater than 1 second, with:
N
A
K
CR
Controller address
Information obtained by the serial interface are listed in
the following table where the column "TYPE" points out
a logic variable with “D” (allowed values "000000" or
"111111") and a numeric variable with "N".
Variables with the first character of "Parameter
Number" equal to X can be either written or read.
Variables with that character equal to “0” can be only
read, equal to "1" can be only written.
Содержание Turbo-V 81 969-9538
Страница 2: ...Turbo V 81 PCB 24 Vdc Controller...
Страница 3: ......
Страница 4: ......
Страница 9: ...ISTRUZIONI PER L USO 3 87 900 121 01 B Cavo interconnessione tra Controller e pompa...
Страница 12: ...GEBRAUCHSANLEITUNG 6 87 900 121 01 B Verbindungskabel zwischen Controller und Pumpe...
Страница 15: ...MODE D EMPLOI 9 87 900 121 01 B C ble d interconnexion entre le contr leur et la pompe...
Страница 18: ...INSTRUCCIONES DE USO 12 87 900 121 01 B Cable de interconexi n entre controlador y bomba...
Страница 21: ...INSTRU ES PARA O USO 15 87 900 121 01 B Cabo de Interconex o entre o Controller e a Bomba...
Страница 24: ...GEBRUIKSAANWIJZINGEN 18 87 900 121 01 B Verbindingskabel tussen controller en pomp...
Страница 27: ...BRUGSANVISNING 21 87 900 121 01 B Forbindelseskabel mellem styreenheden og pumpen...
Страница 30: ...BRUKSANVISNING 24 87 900 121 01 B Anslutningskabel mellan styrenhet och pump...
Страница 33: ...BRUKERVEILEDNING 27 87 900 121 01 B Forbindelseskabel mellom styreenheten og pumpen...
Страница 36: ...K YTT OHJEET 30 87 900 121 01 B Valvojan ja pumpun v linen liitoskaapeli...
Страница 39: ...ODHGIES CRHSEWS 33 87 900 121 01 B Empro sqioj pi nakaj tou Hand Held Terminal...
Страница 42: ...HASZN LATI UTAS T S 36 87 900 121 01 B Vez rl s szivatty k z tti sszek t k bel...
Страница 45: ...INSTRUKCJA UZYTKOWANIA 39 87 900 121 01 B Kabel laczacy kontroler z pompa...
Страница 48: ...P RU KA K POU IT 42 87 900 121 01 B Propojovac kabel mezi Kontrol rem a erpadlem...
Страница 51: ...N VOD K POU ITIU 45 87 900 121 01 B K bel prepojenia medzi Controllerom a erpadlom...
Страница 54: ...NAVODILA ZA UOPRABO 48 87 900 121 01 B Interkonektorski kabel med Controllerjem in rpalko...
Страница 57: ...INSTRUCTIONS FOR USE 51 87 900 121 01 B Controller Pump Interconnection Cable...