TECHNICAL INFORMATION
60
87-900-121-01 (
B
)
Serial Communication Windows
WIN
TYPE
R W
Description
000
L
X
X
START/STOP (1= START ; 0= STOP)
001
L
X
X
SPEED SELECTION ACTIVE [0=HS / 1=LS]
008
L
X
X
REMOTE/SERIAL Configuration (1= Remote ; 0= Serial)
100
L
X
X
SOFT START YES/NO (1= YES ; 0= NO)
Default = 0
101
N
X
X
SET POINT TYPE (0 =Frequency; 1=Current; 2=Time) Default = 0
102
N
X
X
SET POINT VALUE [Hz, Ma, SEC] Default = 1117 Hz
103
N
X
X
SET POINT MASK TIME [s] Default = 0
104
L
X
X
OUTPUT STATE (0= OFF if Val > threshold; 1 = ON if Val > threshold Default = 0
105
N
X
X
HYSTERESIS IN THRESHOLD % [0-100] Default = 2
107
L
X
X
ACTIVE STOP (0=NO; 1=YES)
Default = 0
108
N
X
X
BAUD RATE (3-4) [4800-9600]
Default = 4
109
L
X
PUMP LIFE RESET [Write “1” to Reset]
111
N
X
X
SET ANALOG OUTPUT TYPE [0=Power; 1=Current] Default = 0
120
N
X
X
SET ROTATIONAL FREQUENCY [Hz] 1100 Hz < = F_imp < = FMAX
121
N
X
X
MAX SETTABLE ROTATIONAL FREQUENCY [Hz] F<=F_MAX_ABS
127
N
X
RUN UP TIME (8 – 60 minute)
200
N
X
CURRENT [mA]
201
N
X
VOLTAGE[V]
202
N
X
POWER [W]
203
N
X
DRIVING FREQUENCY [Hz]
204
N
X
PUMP TEMPERATURE [°C]
205
N
X
STATUS [0=stop; 1=interlock; 2=ramp; 3=regulation; 4=brake; 5=normal; 6=failure]
206
N
X
ERROR CODE:
Too high load
No connection
Short circuit
Pump overtemp
SoftStart Not Ended
Controller overtemp
Normal Not Reached
Power fail
211
N
X
CONTROLLER SENSOR TEMPERATURE READING [208= 25°C - 128= 60°C]
216
N
X
AMBIENT SENSOR TEMPERATURE READING
300
N
X
CYCLE TIME [min]
301
N
X
CYCLE NUMBER
302
N
X
PUMP LIFE [h]
319
A
X
Controller Model
320
A
X
Base Pump Model Number (8 characters)
321
A
X
Modified Standard Model Number (4 characters)
323
A
X
Controller Serial Number (5 characters)
325
A
X
Electrical Modification Level (10 characters)
400
A
X
CRC PROGRAM LISTING [QE7xxxx]
401
A
X
CRC BOOTLOADER [BL1xxxx]
402
A
X
CRC PARAMETER LISTING [PA7xxxx]
404
A
X
CRC FILE PARAMETER STRUCTURE
406
A
X
PROGRAM LISTING CODE & REVISION
407
A
X
PARAMETER LISTING CODE & REVISION
500
L
X
MONITOR MODE
WIN
= Window
L
= Logical
R
=
Read
N
= Numeric
W
=
Write
A
= Alphanumeric
7
0
Содержание Turbo-V 81 969-9538
Страница 2: ...Turbo V 81 PCB 24 Vdc Controller...
Страница 3: ......
Страница 4: ......
Страница 9: ...ISTRUZIONI PER L USO 3 87 900 121 01 B Cavo interconnessione tra Controller e pompa...
Страница 12: ...GEBRAUCHSANLEITUNG 6 87 900 121 01 B Verbindungskabel zwischen Controller und Pumpe...
Страница 15: ...MODE D EMPLOI 9 87 900 121 01 B C ble d interconnexion entre le contr leur et la pompe...
Страница 18: ...INSTRUCCIONES DE USO 12 87 900 121 01 B Cable de interconexi n entre controlador y bomba...
Страница 21: ...INSTRU ES PARA O USO 15 87 900 121 01 B Cabo de Interconex o entre o Controller e a Bomba...
Страница 24: ...GEBRUIKSAANWIJZINGEN 18 87 900 121 01 B Verbindingskabel tussen controller en pomp...
Страница 27: ...BRUGSANVISNING 21 87 900 121 01 B Forbindelseskabel mellem styreenheden og pumpen...
Страница 30: ...BRUKSANVISNING 24 87 900 121 01 B Anslutningskabel mellan styrenhet och pump...
Страница 33: ...BRUKERVEILEDNING 27 87 900 121 01 B Forbindelseskabel mellom styreenheten og pumpen...
Страница 36: ...K YTT OHJEET 30 87 900 121 01 B Valvojan ja pumpun v linen liitoskaapeli...
Страница 39: ...ODHGIES CRHSEWS 33 87 900 121 01 B Empro sqioj pi nakaj tou Hand Held Terminal...
Страница 42: ...HASZN LATI UTAS T S 36 87 900 121 01 B Vez rl s szivatty k z tti sszek t k bel...
Страница 45: ...INSTRUKCJA UZYTKOWANIA 39 87 900 121 01 B Kabel laczacy kontroler z pompa...
Страница 48: ...P RU KA K POU IT 42 87 900 121 01 B Propojovac kabel mezi Kontrol rem a erpadlem...
Страница 51: ...N VOD K POU ITIU 45 87 900 121 01 B K bel prepojenia medzi Controllerom a erpadlom...
Страница 54: ...NAVODILA ZA UOPRABO 48 87 900 121 01 B Interkonektorski kabel med Controllerjem in rpalko...
Страница 57: ...INSTRUCTIONS FOR USE 51 87 900 121 01 B Controller Pump Interconnection Cable...