Models 947, 948 and 960 Leak Detectors Operations and Service Manual
2-11
If the I/O on J120 is used, then:
❑
+5 VDC to +24 VDC must be applied to J120, pins 27 and 28.
❑
The +5 VDC to +24 VDC return must be connected to J120, pins 1 and 25.
If the I/O on J121 is used, then:
❑
+5 VDC to +24 VDC must be applied to J121, pins 20 and 21.
❑
The +24 VDC return must be connected to J121, pins 1, 2, 24, and 25.
❑
The grounds between J120, J121, and chassis grounds are exclusive.
When using the discrete I/O, Allen-Bradley’s 1746-IBx and 1746-OBx (or equivalent)
modules for the SLC 500-series PLCs are recommended.
The analog leak rate output voltage on J121, pin 22, and return on J121, pin 23, have the
return referenced to the system ground.
The discrete inputs are:
❑
Opto-isolated with the negative side of the inputs combined and tied to the return
pins of the associated connector.
❑
The input voltage range is +5 VDC to +24 VDC.
❑
The maximum input current (at 24 VDC) drawn by each input is 7 mA.
The discrete outputs are:
❑
Opto-isolated Darlington transistor emitter followers with the collectors combined
and tied to the +24 VDC pins of the associated connector.
❑
The maximum current that can be drawn from each output is 20 mA.
❑
The output voltage supply range is +5 VDC to +24 VDC.
NOTE
When operating the 947/948/960 from the discrete I/O, it is
necessary to make the PARALLEL ENABLE (J120 pin 6) active.
This allows the controller to accept commands from the other
discrete I/O inputs. Making PARALLEL ENABLE active also
disallows any changes to the operating parameters through the
front panel or serial ports (service pendant). Requests for
changes made through the serial ports are responded to with
cant
.
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......