Models 947, 948 and 960 Leak Detectors Operations and Service Manual
B-24
B.5.1.2 Replacing the Wet System Forepump
Before you replace the forepump, examine and clean all O-rings, centering rings, clamps,
and manifold.
CAUTION
There may be oil on these parts. Take care not to place them
where other parts may be contaminated.
B.5.1.2.1 Examining and Cleaning Centering Rings, O-rings and Manifold
Refer to “Examining and Cleaning Centering Rings, O-rings and Manifold” on page B-12.
Execute steps 1 through 4.
B.5.1.2.2 Re-installing the Wet System Forepump
WARNING
To avoid injury, use proper lifting techniques when moving
pumps. Your system may have pumps that require two people
to move them safely.
To perform this procedure:
1. Ensure the area is clean for the new pump.
2. Place the new pump in the 947/948/960 oriented as shown in Figure B-10 on
page B-15, placing the mounting bolts into the mounting holes in the bottom of the
cabinet.
3. Place the lower section of the foreline into the tygon hose section to meet the upper
foreline section.
4. Place the centering ring and O-ring against the pump outlet on the left side of the pump.
5. Place the lower section of the foreline against the centering ring.
6. If necessary, shift the pump so that the connection is clean and there is no strain on the
foreline.
7. Place the NW25 clamp onto the connection and tighten fully.
8. Retighten the radiator clamp around the tygon hose.
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......