Models 947, 948 and 960 Leak Detectors Operations and Service Manual
4-22
3. Hold the spectrometer tube against the centering ring, attach the NW25 clamp, and
tighten the clamp fully.
4. Reconnect the ion source connector, the preamplifier locking D-sub connector, the
Molex connector to the heater, and the suppressor.
a. When attaching the locking connector (Figure 4-9), press the tabs on the sides of the
connector
b. Push the connector onto its mating part.
c. Slide the tabs to lock the connector in place.
Figure 4-9
947/948/960 Spectrometer Tube Installation
NOTE
To ensure reliable testing, after servicing any part of the
vacuum system, the 947/948/960 must be leak checked again.
Refer to “Leak Checking” on page 4-5 for guidelines.
5. Replace the front panel and connect mains power.
6. Turn the leak detector on and wait for the system to be ready.
Allow the pumps to warm up if the detector has been off for some time.
7.
Models 948/960 only:
Once the system has pumped down, clean and fill the Liquid
Nitrogen Cold Trap using the procedure described in “Liquid Nitrogen Cold Trap
(Models 948/960 only)” on page 4-10.
Follow all warnings and cautions in the procedure.
8. Calibrate and tune the 947/948/960 using the procedure described in “Re-calibrate the
Internal Calibrated Leak” on page 4-7.
9. Leak check the assembly before using the 947/948/960 for testing.
Heater Power
Ion Source Connector
Suppressor
Preamplifier Locking Connector
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......