Models 947, 948 and 960 Leak Detectors Operations and Service Manual
3-2
During operation, the LED functions are illuminated to indicate the operation being
performed. The definitions of the operations are as follows:
VENT
Illuminates when the vent valve is open. The LED is off when the
valve is closed.
START
Illuminates when the START button is pushed. Goes off one
second after the TEST LED illuminates, or immediately when the
VENT, HOLD, or NOT READY LEDs illuminate.
TEST
Illuminates when 947/948/960 is in test mode; either the gross leak
valve or the test valve is open. Otherwise, the LED is off.
ZERO
Illuminates when the ZERO button is pushed or when the auto-zero
sequence is performed as part of the auto-calibrate sequence. Goes
off one second after the auto-zero sequence is completed or if the
leak detector goes into a not-ready condition. The ZERO LED
flashes when the signal falls below zero. The ZERO LED also
flashes during startup when the system is nulling the preamplifier.
REJECT
Illuminates when the reject function is on and when the leak rate
reading exceeds the selected set point. Goes off when the leak rate
falls below the set point.
HOLD
Illuminates if the machine is in the HOLD mode with the test and
roughing valves closed. The HOLD mode is selected by momen-
tarily pressing the VENT button if the machine is in TEST. When in
HOLD, pressing either the START or VENT button launches the
next operation.
NOT READY
Illuminates whenever a not-ready condition exists. The particular
error condition code, c0 through c9, is displayed on the RANGE
DISPLAY. See Table 3-1 on page 3-11 for information on the error
codes.
AUTO CAL
If the leak detector is in TEST mode, illuminates when the AUTO
CAL button is pushed, disabling all other front panel push buttons,
including START and VENT. AUTO CAL is only functional when
the leak detector is in TEST mode. The machine returns to TEST
mode at completion of the AUTO CAL routine.
STD LEAK
Illuminates when the standard leak test valve is opened. This
occurs when the STD LEAK button is pushed or during the auto-cal-
ibrate function. Goes off when the standard leak test valve is closed
by pressing the STD LEAK button.
SERVICE
Not used.
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......