Models 947, 948 and 960 Leak Detectors Operations and Service Manual
B-25
9. Place the centering ring and O-ring against the pump outlet on the right side.
10. Place the exhaust fitting against the centering ring.
11. Place the NW16 clamp onto the connection and tighten fully.
12. Working from under the leak detector, place the nuts onto the mounting bolts and
tighten them fully.
13. Connect the pump power cord to the 947/948/960.
14. Consult the pump operations manual for any other necessary procedures to be
performed before closing the panels or powering up the pump
CAUTION
To ensure reliable testing, after servicing any part of the
vacuum system, the 947/948/960 must be leak checked again.
Refer to “Leak Checking” on page 4-5 for guidelines.
15. Replace the front panel.
16. Replace the side panel.
17. Reconnect the mains power supply.
18. Turn the leak detector on and wait for the system to be ready.
Allow the pumps to warm up if the detector has been off for some time.
19.
Models 948/960 only:
Once the system has pumped down, clean and fill the Liquid
Nitrogen Cold Trap using the procedure described in “Liquid Nitrogen Cold Trap
(Models 948/960 only)” on page 4-10.
Follow all warnings and cautions in the procedure.
20. Calibrate the 947/948/960 using the procedure described in “Re-calibrate the Internal
21. Leak check the assembly before using the 947/948/960 for testing.
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......