7. Place the foreline against the turbo flange centering ring and O-ring.
NOTE
Gently pivot the turbo pump until it is in the position to make
the connection without placing stress on the manifold. The
claw clamps should be loose enough to allow you to do this.
8. Place the NW16 clamp onto the connection and tighten the wing nut.
9. Fully tighten the ISO single claw clamps.
10. Connect the turbo pump power by pushing the power connector onto the pump and
then turning the green sleeve until the connector locks into place.
11. Connect the mechanical pump power plugs if they are not connected already.
12. Place the turbo fan assembly against the turbo pump, oriented so that the fan is on the
front of the pump with the guard facing outward.
13. Place the radiator clamp around the turbo pump. Tighten the radiator clamp screw fully.
CAUTION
To ensure reliable testing, after servicing any part of the
vacuum system, the 947/948/960 must be leak checked again.
Refer to “Leak Checking” on page 4-5 for guidelines.
14. Replace the side panel.
15. If the front panel was removed, replace it.
16. Reconnect the mains power supply.
17. Turn the leak detector on and wait for the system to be ready.
Allow the pumps to warm up if the detector has been off for some time.
18. Calibrate the 947/948/960 using the procedure described in “Re-calibrate the Internal
19. Leak check the assembly before using the 947/948/960 for testing.
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......