AutoSamplers
Page: 112
User Manual 450-GC
Varian, Inc.
CALIBRATION
Sampler calibration will align the Sampler components (Tower, Vials and
Needle) and the GC Injector ports.
Calibration procedure is important for a trouble free working GC/Sampler
combination.
Calibration must be carried out in the following cases:
•
First Installation of the Sampler.
•
Sampler has been removed from the GC and after maintenance/repair back
on the GC.
•
Bending Sample needles.
•
Other Injector positions used (back or front).
•
If the Injector has been removed calibration is mandatory.
Do not recalibrate the Injector positions after normal Injector maintenance such
as changing septa or inserts.
Two calibration methods are present:
- Vial calibration, on page 113.
- GC Injector calibration, on page 116.
If during Calibration a question mark appears in a yellow triangle,
this wil l indicate that the calibration is not corr ect!