Suggested PFPD Adjustment Settings
The following adjustment settings are recommended for setup and routine
operation of the PFPD. These adjustments are accessed by pressing the
Adjustments button in the PFPD detector section of the 450-GC method.
Element
Gate Delay
Gate Width
Trigger
Sulfur
6
20
200
Phosphorus
4
10
200
Nitrogen
3.5
8
200
Optimization Criteria
Refer to the following table when optimizing the PFPD.