1177 Split/Splitless Capillary Injector - Injectors
Varian, Inc.
User Manual 450-GC
Page: 169
Testing the 1177 Injector Performance
The following procedure describes how to test the performance of the 1177
Capillary Injector. This is best done with a test sample containing an
appropriate set of components for the installed detector. The following table
lists the series of test samples available for 450-GC.
Test Sample
Part Number
Concentrati ons of Test Compoun ds
TCD
82-005048-01
3.00
μ
g/
μ
L of C
14
, C
15
and C
16
in iso-octane.
ECD
82-005048-02
33.0 pg/
μ
L of lindane and aldrin in iso-octane.
PFPD
82-005048-03
20.0 ng/
μ
L of n-dodecanethiol, tributylphosphate,
methyl parathion; 4000 ng/
μ
L of n-pentadecane in
iso-octane.
TSD
82-005048-04
2.00 ng/
μ
L of azobenzene, methyl parathion; 4.00
ng/
μ
L malathion and 4.00
μ
g/
μ
L C
17
in iso-octane.
FID
82-005048-07
30 ng/
μ
L of C
14
, C
15
and C
16
in iso-octane.
Note:
If the FID test sample is not available, the TCD test sample can be used if first
diluted 100:1.
Table 31: Detector Test Samples
To run one of these test samples, use the chromatographic conditions listed
previously for the injection technique you are currently using. The detector
should be operated at the most sensitive range, e.g., 12 for FID and TSD, 0.05
for TCD, 10 for PFPD and 1 for ECD. The resultant chromatogram on page 419
should approximate that shown in the detector section of this manual.
Содержание 450-GC
Страница 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Страница 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Страница 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Страница 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Страница 192: ...Replacement Parts Replacement Parts Page Page 456 456 User User Manual Manual 450 GC 450 GC Varian Varian Inc Inc...
Страница 201: ......