AutoSamplers
Page: 118
User Manual 450-GC
Varian, Inc.
5. If the needle guide descends so that it is either resting on the inject switch
or below the needle cone on the inject switch, use the “Up” button move the
sled up (10 steps) until the needle guide is just above the Injector cone.
6. Using the “CCW” (counterclockwise) and “CW” (clockwise) button. Position
the tower directly over the Injector Inject Switch Locator cone.
7. Adjust the height (using the “Up” and “Down” button) of the needle guide
such that the sled is just resting on or very slightly depressing the Injector
Switch and is covering the locator cone.
8. Press the “Accept” button if the calibration of Injector 1 is complete, the
calibration data will be saved.
9. Calibrate Injector 2 (if present) using the above mentioned procedure.
10. Press the “Exit” button if calibration is finished or leaving the calibration
page without saving is needed.
The “ Reset Calibration” button wi ll restore the factory settings to all
calibration values.