AutoSamplers
Page: 146
User Manual 450-GC
Varian, Inc.
Injector Temperature
To accelerate the desorption process of the analytes from the SPME fiber,
elevated injection temperatures are required. The ideal temperature is
approximately 10-20°C below the maximum temperature of the fiber and GC
column (usually between 200° and 280°C).
Sample Vials and Septa
SPME is a concentration technique normally used in combination with a
GC/MS. To minimize pollution of the analysis by the vial septum, special Ultra
clean septa are used. These vials are included in the SPME kit. Replacement
parts see here, on page 441.
Agitation Strokes
SPME can be used in the agitation “mode”, which is useful when performing
liquid SPME. Agitation disrupts the depletion zone that may surround the fiber
during the extraction process. With CP-8400 and CP-8410 SPME, agitation is
achieved by moving the fiber in and out of the sample liquid, approximately
once every 2 seconds.
To use the agitation “mode”, fill in the number of Agitation Strokes (1 or more).
Zero will disable the Agitation “mode”.