AutoSamplers
Page: 116
User Manual 450-GC
Varian, Inc.
GC Injector calibration
1. Press the SETUP
key and select CALIBRATION, select on the next
page Sampler CALIBRATE.
2. The calibration screen will be visible.
3. Before starting the calibration identify Injector 1 and Injector 2.
Injector 1 is the first Injector the Tower will encounter when it rotates
clockwise to reach the Injector(s). The second Injector is Injector 2. Please
notice the difference when the sampler is mounted in the Front or Rear
(next page) position.
Injector 2
Injector 1
Sampler mounted in
the Front position
Содержание 450-GC
Страница 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Страница 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Страница 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Страница 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Страница 192: ...Replacement Parts Replacement Parts Page Page 456 456 User User Manual Manual 450 GC 450 GC Varian Varian Inc Inc...
Страница 201: ......