Injectors – 1177 Split/Splitless Capillary Injector
Page: 156
User Manual 450-GC
Varian, Inc.
1177 Injector Screen
The 1177 Injector screen allows modifying the Injector parameters.
With the Enable checkbox it is possible to enable/disable the Injector
temperature control.
In the Temperature box the wanted Injector temperature can be set.
The Split State controls the flow of carrier gas through the Injector during the
analytical run. When the Split State is unchecked, most of the sample injected
is directed onto the column. When the Split State is checked, the sample is split
in the Injector with typically the smaller fraction entering the column and the
larger fraction being vented.
In the example below is a splitless injection. The Injector is held in the split state
for the initial period with split ratio of 20. When the run is started the split state
will switched Off. After 0.9 minute the split is switched on again.
Содержание 450-GC
Страница 92: ...AutoSamplers Page 152 User Manual 450 GC Varian Inc...
Страница 144: ...Detectors Flame Ionization Detector FID Page 294 User Manual 450 GC Varian Inc...
Страница 177: ...Maintenance Page 438 User Manual 450 GC Varian Inc...
Страница 179: ...Shipping Cleaning and Disposal Instructions Page 440 User Manual 450 GC Varian Inc...
Страница 192: ...Replacement Parts Replacement Parts Page Page 456 456 User User Manual Manual 450 GC 450 GC Varian Varian Inc Inc...
Страница 201: ......