AutoSamplers
Varian, Inc.
User Manual 450-GC
Page: 105
INSTALLING THE CP-8400/8410
If the Sampler has been installed by a service engineer, you can skip this
chapter and goto page: 122.
Before installing the Sampler, turn off the oven and any heated
zones and let them cool down. Turn off any detector g ases at their
supply. Turn off t he main power swi tch and unplug th e power cord.
Use the correct (included) screws (metric) to mount t he sampler.
1. Before installing the CP-8400 or CP-8410 AutoSampler, make certain your
450-GC is equipped with the appropriate Top-Cover. This cover will have a
long rectangular opening for the installation of the Injector locator plate,
markings indicating the front and rear installation positions for the
AutoSampler and a small rectangular opening to route the CP-8400/8410
power/communication cable through.
Figure 14: Top-Cover for Sampler