User’s Guide ______________________________________________________________________
30 ___________________________________________________________________M211486EN-A
Ozone Destruction Filter
The ozone destruction filter eliminates ozone generated during testing.
lists the parts required for the ozone destruction
filter. However, the ASOPOS panel (Assessment of Standard Operating
Procedures for Ozonesondes) recommends the use of purified, ozone-
free air instead of the destruction filter. Refer to the GAW report 201 for
more information. See Appendix D
Performance Review Literature on
.
Assemble the ozone destruction filter components as shown in
.
Table 6
Equipment Required for Ozone Destruction
Filter
Item Required
Vaisala Part
Number
Particle and ozone filter: Mine Safety Appliances (MSA)
Company. Delivered by MSA as a single cartridge (part
number 815185).
234561 (1 pc)
Funnel, 75 mm in diameter, tube diameter 8 mm, glass
12725
Connector tube 1: soft silicon tube approximately 5 cm
long, inner diameter 6 mm, outer diameter 10 mm
12642OS (set of
tubes)
Connector tube 2: soft vinyl tube approximately 60 cm
long, I.D. 1/8" (3.2 mm), O.D. 1/4" (6.4 mm)
12642OS (set of
tubes)
Connector tube 3: soft silicon tube approximately 2 cm
long, inner diameter 2 mm, outer diameter 4 mm
12642OS (set of
tubes)
Connector tube 4: Cut a piece of tubing from the ECC
sensor air intake tube (2 cm), or order separately.
17348S (by the
meter) or SPC
spare part No.
OTU-19
Electrical tape (Nitto 15)
4103