Arrange the air flow meter as shown in
procedure for air flow measurement is described below. The numbers
within brackets refer to the items in
:
1.
Fill the rubber bulb (4) and the flow meter tube with soap solution
(8) almost up to the filling tube of the flow meter tube.
2.
Connect the air flow meter to the sensor cathode air exhaust tube.
This is done by slipping the connector tube 3 (9) over the short
Teflon tube protruding from the top plug of the sensor cathode
chamber.
3.
With the radiosonde air pump operating, squeeze the rubber bulb
(4) slightly to cause several soap bubbles to rise up the flow meter
tube. Repeat the process several times, until the bubbles reach the
top of the tube without breaking.
4.
Now form only one bubble, and use a stop-watch to determine the
time (t) required for the bubble to rise from 0 to 100 ml up the flow
meter tube (5). Repeat the measurement three times to obtain a
mean value.
NOTE
When the air flow is measured, make sure the sensor is charged with
the sensing solution.