Chapter 2 ______________________________________ Product and System Component Overview
VAISALA_______________________________________________________________________ 35
Air Flow Meter
lists the equipment required for the Air Flow
Meter. Section
Using the Air Flow Meter on page 35
explains how the
air flow meter is used in ozone sounding preparations.
shows an example of an air flow meter.
Using the Air Flow Meter
This procedure is meant to be carried out with no remarkable breaks
(that is, breaks lasting over two hours).
Table 10
Equipment Required for Air Flow Meter
Item Required
Vaisala Part
Number
Air flow meter tube. Burette with filling tube, capacity
100 ml
12733
Rubber bulb, capacity approximately from 50 to 80 ml
12734
Burette stand with two bossheads and two clamps
12730 (stand),
12732 (bosshead),
12728 (clamp)
Stop-watch, accuracy at least 0.1 s
12784
Connector tube 1: soft silicon tube approximately 5 cm
long, inner diameter 6 mm, outer diameter 10 mm
12642OS (set of
tubes)
Connector tube 2: soft vinyl tube approximately 60 cm
long, I.D. 1/8" (3.2 mm), O.D. 1/4" (6.4 mm)
12642OS (set of
tubes)
Connector tube 3: soft silicon tube approximately 2 cm
long, inner diameter 2 mm, outer diameter 4 mm
12642OS (set of
tubes)
Dishwashing liquid: Add about one teaspoon of
dishwashing liquid and one teaspoon of glycerol to 1 dl
of water.
NOTE
To avoid contamination, note the following precautions:
Work in a clean environment with clean hands.
Never operate the pump without the ozone destruction filter or purified
air.
Do not use a sensor loaded with solutions if the sensor is not connected
to a powered interface (or if the anode and cathode wires are
connected).
Содержание RS41
Страница 1: ...USER S GUIDE Ozone Sounding with Vaisala Radiosonde RS41 M211486EN A ...
Страница 124: ...www vaisala com M211486EN ...