User’s Guide ______________________________________________________________________
44 ___________________________________________________________________M211486EN-A
Pump voltage:____________ V DC
Pump current:____________ mA
Flow rate:_____________s/100 ml
3.
Connect the ozone sensor to the ozonizer (motor, output, sensor
leads).
a.
Turn on the ozonizer.
b.
Record I
B0
before applying any ozone to the sensor:
I
B0
_____________
µ
A
4.
Condition the pump and the dry sensor with HI O
3
for 30 minutes.
5.
After 10 minutes of HI O
3
, check the following values from the
ozonizer:
(Normal head pressure and vacuum values are 700 ... 900 hPa)
6.
Turn off HI O
3
. Run NO O
3
for 5 minutes.
7.
Next, you must charge sensor cathode and anode. Use a different
syringe for cathode and anode.
Charge sensor
cathode
with solution 3.0 cm
3
and wait 2 minutes.
Charge sensor
anode
with solution 1.5 cm
3
.
8.
The current starts to decrease. Sensor background current after 10
minutes (typically the value is under 1.5
µ
A) on NO O
3
:
____________
µ
A
9.
Run on moderate ozone (about 5
µ
A) for 10 minutes.
Measured
Limit values
Pump
voltage
_________ V
12 ... 13 V
Pump
current
_________ mA
SPC: < 115 mA; DTM: <100 mA
Head
pressure
_________ Pa
>
670 hPa app. 20 in Hg
Vacuum
(this is a
minus
value)
_________ Pa
>
670 hPa app. 20 in Hg