Chapter 3 _________________________________________________________ Ozone Calculation
VAISALA_______________________________________________________________________ 89
where:
The partial pressure of ozone (P
3
) is:
R = 8.31451 JK
-1
mol
-1
(Molar gas constant)
Finally:
where
C
=
ozone concentration in mmol 1
-1
F
=
9.6485
×
10
4
C (mol)
-1
(Faraday constant)
I
=
measured current in
µ
A
t
=
pumping time for 100 ml of air (seconds)
P
3
=
partial pressure of ozone in mPa
I
=
measured ozone current in
µ
A
I
BG
=
current caused by oxidants other than ozone (mainly O
2
) in
µ
A (I
BG
equals to I
B2
).
Tp
=
measured airflow temperature in K from pump base.
t
=
pumping time for 100 ml of air in seconds
C
ef
=
correction due to reduced ambient pressure for pump
C
ref
=
additional correction factor
NOTE
I
BG
used in this manual is equal to I
0
used in the Vaisala scripts, and
I
B2
used in the GAW report 201, and the ASOPOS panel
recommendations.
P
3
C R T
air
R
F
2 100
ml
-------------------------------
I T
air
t
=
=
P
3
4.3087 10
4
–
I I
BG
–
T
p
t C
ef
C
ref
=
Содержание RS41