User’s Guide ______________________________________________________________________
66 ___________________________________________________________________M211486EN-A
Preparing the Radiosonde with
Ground Equipment
1.
Start MW41 sounding software, if you have not started it yet, and
log in.
2.
Attach the ozone destruction filter to the pump inlet tube.
3.
Place the RS41 radiosonde on the ground check device. The
radiosonde is switched on when you place it on the ground check
device. The message "Preparation in progress" will be displayed in
MW41.
4.
Before the preparation phase is completed, scroll the MW41 page
down to the Special sensor window, and select
Ozone
from the
drop-down list.
5.
Fill in the information needed and click
Apply
.
During this phase, the radiosonde LED light is red, but you can
ignore it. It does not indicate an error at this point.
CAUTION
To be able to prepare an ozone sounding with MW41, you must import
and activate the following scripts:
OzoneCalculations.py
,
OzoneMain.py
, and
OIF411.py
. The scripts are included on the
MW41 installation DVD, in folder ScriptLibrary\CalcOzone. The
ozone sounding cannot be completed if any of these scripts is missing.
Import the scripts to the same Script Group and select OzoneMain.py
as the main script. Make sure that Script group is set active. You do not
need to set a command line argument. For more instructions, see
MW41 on-line help.
CAUTION
Do not connect the Ozone Interface Board OIF411 to the radiosonde
while the radiosonde is placed on the ground check device. Connecting
OIF411 during the ground check will interrupt the preparations and
MW41 will return to the Radiosonde selection window.
NOTE
The ASOPOS panel recommends 3.0 cm
3
for cathode solution
volume.