User’s Guide ______________________________________________________________________
92 ___________________________________________________________________M211486EN-A
For other sensor manufacturers, the pump correction is very similar to
the SPC ozone sensor; only the table values differ.
For Model Z, use the following table:
Table 13
Ozone Partial Pressure Correction Factors
Ozone Partial Pressure
Correction Factor C
ef
Atmospheric
pressure hPa
Sensor cathode solution volume
2.5 cm
3
Sensor cathode solution
volume 3.0 cm
3
2.0
1.160
1.171
3.0
1.124
1.131
5.0
1.087
1.092
10.0
1.054
1.055
20.0
1.033
1.032
30.0
1.024
1.022
50.0
1.015
1.015
100.0
1.010
1.011
200.0
1.007
1.008
300.0
1.005
1.006
500.0
1.002
1.004
1000.0
1.000
1.000
Table 14
Ozone Partial Pressure Correction Factors
P/hPa
C
ef
≤
3
1.24
5
1.124
7
1.087
10
1.066
15
1.048
20
1.041
30
1.029
50
1.018
70
1.013
100
1.007
150
1.002
≥
200
1