Chapter 3 _____________________________________________ Constructing an Ozone Sounding
VAISALA_______________________________________________________________________ 77
Always record the value indoors with destruction filter or zero air
connected, and after inserting the ozone sensor to the flight box.
At this stage, it is recommended to power the ozone sensor with an
external power source, not with the ozone sensor battery.
13. MW41 proposes a background current. Use the value by clicking
Copy
. You can either use the
Copy
button or enter the sensor
background current in MW41 and click
Apply
.
1403-192
Figure 49
Entering Sensor Background Current in MW41
14. In the Monitoring window, go to the
Tabular data
tab and select
Raw ozone
from the drop-down list. Make sure that ozone data is
coming through, showing reasonable ozone current and ozone
flight box temperature.
1403-169
Figure 50
Viewing Raw Ozone Data in MW41
The following information is displayed:
-
Time: time in seconds
-
Raw O3: ozone partial pressure
-
Current
-
Box temperature
-
Ozone pump current
15. Turn off the pump and either take off the ozone destruction filter or
shut off zero air flow.