Chapter 3 _____________________________________________ Constructing an Ozone Sounding
VAISALA_______________________________________________________________________ 45
10. Check the sensor response to LO O
3
. It should be about 5
µ
A.
11. Run on NO O
3
for 10 minutes. Record the background current:
I
B1
____________
µ
A
12. Switch everything off and short-circuit the ozone sensor wires.
13. Put the ozone sensor back to the flight box and store it in a dark,
clean-air environment at a temperature of 20 ... 25 °C until it is
used.
After these preparations, a sensor cleaning process takes place, whereby
both half cells (anode and cathode) and ionbridge get in balance.
Содержание RS41
Страница 1: ...USER S GUIDE Ozone Sounding with Vaisala Radiosonde RS41 M211486EN A ...
Страница 124: ...www vaisala com M211486EN ...