SECTION 6
OPERATION
58
6
Table 6-5
Remote-control Signals Setting Mode
(Note 1) When set to EI-03, behavior of remote-control signals is the same as ULVAC Turbo Molecular
Pump power supply "EI-xx03M" and "EI-R04" series.
(Note 2) When set to SEMI E74, behavior of remote-control signals conform to SEMI E74 standard
"Specification for Vacuum Pumps - Turbomolecular Pumps".
Refer to the same standard about the shape of connectors and the pin configuration.
6.7.3 Connector
Fig. 6-4
Arrangement of Remote-control Connector Pin (front view)
Signal
Description
Pin No.
ALARM
(23)-(24) (11)-(24)
EI-03
(Note 1)
Alarm occurred
Open
Close
Power OFF and no alarm
Close
Open
SEMI E74
(Note 2)
Power OFF and active alarm
Close
Open
No alarm
Open
Close
WARNING
(13)-(25) (12)-(25)
EI-03
(Note 1)
Warning occurred
Open
Close
Power OFF and no alarm
Close
Open
SEMI E74
(Note 2)
Power OFF and active warning
Close
Open
No alarm
Open
Close
REMOTE
EI-03
(Note 1)
Even if remote control is possible, "REMOTE" signal is OFF during
power interruption regenerative control.
SEMI E74
(Note 2)
"REMOTE" signal is normally ON when remote control is enabled.
STOP
REMOTE
ONLY
"STOP" signal (open between (2) and (16)) is enabled only when
remote control is enabled.
REMOTE &
RS-XXX
"STOP" signal (open between (2) and (16)) is enabled by priority even
during RS-232C and RS-485 operation. Use this setting such as when
using hardware interlock.
Содержание UTM1600 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 26: ...12 2 2 4 External I F Panel This page is intentionally left blank...
Страница 27: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 32: ...18 3 3 3 Controller This page is intentionally left blank...
Страница 38: ...24 4 4 4 Standards Fulfilled This page is intentionally left blank...
Страница 54: ...40 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 78: ...64 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 79: ...7 7GAS PURGE...
Страница 88: ...74 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 101: ...10 10WARRANTY CLAUSES...
Страница 104: ...90 10 This page is intentionally left blank...
Страница 106: ...Index Index 2 This page is intentionally left blank...