
9.5
Alarm Detection Capabilities
81
9
Table 9-4
If the Alarm Lamp Lights
Alarm
code
Alarm name
Possible cause
Remedy
Sec-
tion
1
11
12
TD COUNTER LIMIT
PF COUNTER LIMIT
The number of high
speed or power failure
touch-downs has
exceeded the
prescribed number.
The touch-down bearing may have
deteriorated. Consult ULVAC or an
approved service company regarding
replacement of the touch-down
bearing.
2
15
POWER FAILURE
Power failure or
reduction in the power
supply voltage.
Wait for the power to be restored. An
unfamiliar sound will be heard a few
minutes after a power failure. This
sound occurs when the rotor contacts
the protective bearing as magnetic
levitation cannot be maintained. This
is not an abnormal sound.
POWER switch was
turned off by mistake.
Wait about 5 seconds before turning
the POWER switch back on. Re-
acceleration is possible after resetting
and start-up.
3
16
TMP:OVERLOAD
Drop in rotation speed
during rotation at rated
speed (increased
internal pump
pressure).
Check that the outlet and inlet
pressures are below the specified
maximum pressures.
Check for leakage.
Check that too much purge gas is not
flowing.
Check that process gas flow rate is
not too high.
4.1
4
19
EI:FAILURE
Power supply unit can
not identify the pump
correctly.
Consult your ULVAC representative.
5
21
24
TMP TEMP/MB CABLE
TMP:OVER TEMP
High pump unit
temperature.
Check that the ambient temperature
around the pump is within the
specified range.
Check that the temperature and flow
rate of the cooling water are within
the specified ranges.
Check that no load in excess of the
specified range is continuously
applied to the pump.
4.1
4.3
6
46
47
MOTOR OVERSPEED
EI:R-SPEED ERROR
Rotation detection
signal from the motor
cannot be detected
correctly.
Check that equipment causing noise
is not used around the pump and
power cable.
7
23
34
EI:MOTOR OVERCURR
EI:INV. OVERCURR
Overcurrent supplied to
the motor.
Check that the ambient temperature
around the pump is within the
specified range.
Check that the temperature and flow
rate of the cooling water are within
the specified ranges.
4.1
4.3
5.3
8
30
31
32
EI:CONT.TEMP ERR
EI:BR OVERTEMP
EI:DC-DC OVERTEMP
Temperature increase
in the control system.
Check that the ambient temperature
around the pump is within the
specified range.
Check that the temperature and flow
rate of the cooling water are within
the specified ranges.
4.1
4.3
5.3
Содержание UTM1600 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 26: ...12 2 2 4 External I F Panel This page is intentionally left blank...
Страница 27: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 32: ...18 3 3 3 Controller This page is intentionally left blank...
Страница 38: ...24 4 4 4 Standards Fulfilled This page is intentionally left blank...
Страница 54: ...40 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 78: ...64 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 79: ...7 7GAS PURGE...
Страница 88: ...74 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 101: ...10 10WARRANTY CLAUSES...
Страница 104: ...90 10 This page is intentionally left blank...
Страница 106: ...Index Index 2 This page is intentionally left blank...