SECTION 7
GAS PURGE
66
7
This turbo molecular pump incorporates a gas purge port (Fig. 2-1 (2)).
In case of using a gas purge, the joint can be selected from Table 7-1.
The proper purge gas flow rate is 30 mL/min.
Fig. 7-1 is an example of gas purge piping diagram. Use a filter of 5 μm or less in element size.
Use flow control valve to control the flow rate.
Fig.7-1
Gas Purging Method (adaptor with Φ0.5 mm orifice)
Table 7-1
Table of Gas Purge Adaptor
Start Gas Supply
After starting backing vacuum pump, but before exhausting process gas
Stop Gas Supply
After thoroughly exhausting process gas, but before shutting off backing
vacuum pump
Gas Type
Nitrogen, purity 4N (99.99 %) or better
Joint
Orifice size
Description
KF10 (Clamp / Centering with O-ring)
Φ0.5 mm
GP ADAPTOR, 0.5 1303 KF
KF10 (Dust cap)
Φ0.5 mm
GP ADAPTOR 0.5 1303, KF CAP
UJR 6.35 (Male)
Φ0.5 mm
ADAPTOR, 0.5 UJR
SWAGELOK Φ6.35
Φ0.5 mm
ADAPTOR, 0.5 SWG
4-VCR (Female)
Φ0.5 mm
ADAPTOR, 0.5 VCR
Содержание UTM1600 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 26: ...12 2 2 4 External I F Panel This page is intentionally left blank...
Страница 27: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 32: ...18 3 3 3 Controller This page is intentionally left blank...
Страница 38: ...24 4 4 4 Standards Fulfilled This page is intentionally left blank...
Страница 54: ...40 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 78: ...64 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 79: ...7 7GAS PURGE...
Страница 88: ...74 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 101: ...10 10WARRANTY CLAUSES...
Страница 104: ...90 10 This page is intentionally left blank...
Страница 106: ...Index Index 2 This page is intentionally left blank...