b. Multi-drop function ON
Fig. 6-7
Example of RS-485 Cable Wiring Connections (Multi-drop function ON)
(2) Cables used
RS-485 is a differential transmission and use twisted-pair cables in combinations as shown
in Fig. 6-7. The shield of the cable should be connected to case.
(3) Connecting the terminator
A terminator (120 Ω, 1/4 W min.) is required for connection.
Terminators are not necessary for multidropped turbo molecular pump with another pump
or pump to which the computer connects is connected to both ends (pump 1 through N-1
in Fig. 6-7).
However, connection of the terminator may prevent communications with certain cable
lengths and RS-485 device types. Connect the terminator to determine whether it is
required.
(4) Cable length
Connection cables can be extended up to 1.2 kilometers, but may be subjects to errors
depending on actual operational environment.
Содержание UTM1600 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 26: ...12 2 2 4 External I F Panel This page is intentionally left blank...
Страница 27: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 32: ...18 3 3 3 Controller This page is intentionally left blank...
Страница 38: ...24 4 4 4 Standards Fulfilled This page is intentionally left blank...
Страница 54: ...40 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 78: ...64 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 79: ...7 7GAS PURGE...
Страница 88: ...74 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 101: ...10 10WARRANTY CLAUSES...
Страница 104: ...90 10 This page is intentionally left blank...
Страница 106: ...Index Index 2 This page is intentionally left blank...