Page 36 of 101 pages
4.1.2 Display
The CAT2007 display consists of 4 alphanumeric rows with 20 characters in each row:
First Row
First 3 characters define the Cycle no. (Ranges from 1-14)
There are 8 characters to define the Cycle name.
The final 5 characters at the right of the row are allocated to displaying the status
of the appliance or the stage in which the cycle is found whilst running.
♦
When the autoclave is in waiting stage – ST.BY
♦
When the autoclave is ready to work – READY
♦
When a process ends correctly – END
♦
When a program ends incorrectly – FAIL
The names of the stages follow:
♦
Vacuum
♦
Exh.
(Exhaust)
♦
Heat
♦
Dry
♦
Ster.
(sterilization)
♦
Test
Following are the 14 cycles, names and relevant temperatures:
Temp
Cycle
no.
Program
ºF
01& 02
Unwrapped
250
03 &4
Unwrapped
270/274
05&06
Wrapped
270
07-
Wrapped
274
08-
Wrapped
250
09& 10
Materials that require a slow
pressure exhaust
250
11 & 12-
Materials that require a slow
pressure exhaust
250
13
B&D test
274
14-
Leak test
Second Row:
This row displays the temperature in the chamber, pressure in the chamber, and
jacket pressure.
Third Row:
This row is used to display messages to the operator such as DOOR1 OPEN.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 5596 SP-1
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 12: ...Page 7 of 101 pages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 108: ...AVg Z HiZVb HiZg a oZgh dg i Z A Z HX ZcXZh AVWdgVidgn A cZ...