Page 57 of 101 pages
The temp for 134
º
C needs correction without changing the accurate Temp of
121
º
C.
The data should be set as follows:
Upper Row
A 121.0
R 121.0
A 132.0
R 134.0
The system calibrates the new offset and gain and loads them into the non-
erasable memory.
As a result, while running a 134
º
C program the correction is automatically
made, while at 121
º
C there is no change.