Page 59 of 101 pages
The following is recommended:
Tubing
Rinse tubing after cleaning with pyrogen free water. When placing in tray,
ensure that both ends are open, without sharp bends or twists.
Packs
Place packs upright on trays, side by side.
Packs should not touch the chamber walls.
Sterilization of Liquids
Use only heat-proof glass, filled 2/3 full.
Ensure that the glass container is covered but not sealed to prevent
pressure build-up.
Loading
Every machine is supplied either with two (or more) stainless steel shelves,
or with rails for loading carriage from a loading cart..
The loading cart is made of stainless steel and comes with adjustable legs to
balance the height of the loading cart to rails in the sterilizer chamber.. It is
designed to slide smoothly off the carriage into the sterilization chamber.
The cart moves on four heavy - duty castors, which are mounted on its base.
When loading the materials to be sterilized into the chamber, observe the
rules described in this chapter.
Do not overload the carriage or the shelves as this can have adverse effects
on the results of the sterilization and drying. Leave space between the packs
or wrapped instruments to allow complete removal of air pockets in the air
removal stage, and for free penetration of steam in the heating and
sterilization stages.
chamber.
Transfer Carriage
Loading Cart
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 5596 SP-1
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 12: ...Page 7 of 101 pages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 108: ...AVg Z HiZVb HiZg a oZgh dg i Z A Z HX ZcXZh AVWdgVidgn A cZ...