CALIBRATING TEMPERATURE and PRESSURE
Return to the ST.BY screen and select the OPTION command keypad.
The word CALIB is displayed above the middle command key. Press the
middle keypad and the same screen for entering the SETUP code will appear.
Enter the correct code via the digit keypad. The system begins displaying the
calibration options.
Note:
♦
If the code is not entered within 15 seconds, the display returns to the
Stand BY screen.
♦
If the code is incorrect it will return to the ST.BY screen
♦
In all screens following the code entry screen, the middle command
keypad is Clear and the right keypad is Enter. Pressing on either of these
keys opens the next screen.
♦
The values in all the following screens can only be changed via the
UP/DN keypads. When the correct value is displayed, press Enter and the
system will progress to the next screen.
The following screens appear the entry code screen for Calibration the
following screens appear:
5.1.1 Temp 1:
Temperature of sensor 1.
This screen requires a code Press the UP key once to enter the code value
ØØ1 and press ENTER.
A screen displaying 4 Temp values appears:
A100
R100
A130
R130
Pressing the right button (F3) allows you to move around the different values,
which may be changed using the UP/DN key only.
After F3 is pressed, when it is on the last value, the next screen Will appear.
Pressing the middle button (F2) will read the value of the sensor.
Pressing STOP will stop the cycle.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 5596 SP-1
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 12: ...Page 7 of 101 pages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 108: ...AVg Z HiZVb HiZg a oZgh dg i Z A Z HX ZcXZh AVWdgVidgn A cZ...