Page 64 of 101 pages
8.
OPERATING INSTRUCTIONS
1. Check that the feed water valve is open.
2. Check that the compressed air valve is open.
3. Power the autoclave by turning on the main power switch, observe the
gauge and see that the generator reached its pressure.
4. Check that the jacket pressure gauge, reaches 2.4 bars,
The sterilization cycle can be started
.
5. Open the door by turning the door wheel handle, counter clockwise to
release the locking arms.
6. Load the sterilizer.
7. Close the door by pushing it towards the chamber frame and turning the
wheel handle clockwise until the word READY appears in the first row,
indicating autoclave is ready to start operation.
8. Select the sterilization program according to the type of material to be
sterilized by pressing the relevant keys as described in Chapter 4.
Press the START key, to operate the cycle. The entire sterilization
process automatically proceeds. The signal light RUN on the control
panel is lit.
9. At the end of the cycle a buzzer sounds, and the word END is displayed.
Open the door as described in par 5 while OPEN is displayed above this
key.
10.
Take out the sterilized material from the chamber following the steps
indicated in the ‘Unloading’ paragraph.
Note
: In case of failed cycle or manual stop by operator, the two LED's RUN
and FAIL are lit. The reason for failure is displayed on the screen.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 5596 SP-1
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 12: ...Page 7 of 101 pages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 108: ...AVg Z HiZVb HiZg a oZgh dg i Z A Z HX ZcXZh AVWdgVidgn A cZ...