4.3.4 Changing Cycle Parameters
The screen that allows for parametric changes appears as follows:
Pressing Clear or Enter before entering the code returns to the Stand by screen,
and the RUN & FAIL LED's are lit.
Enter the correct code only via the UP/DN keypads; press ENTER to proceed to
the next screen.
Pressing the UP/DN keypads can also change all parameter values.
Code 1 or 11 Parameters
STE Temp – Required sterilization temperature for the process
Entry code
1 or 11
Resolution
1
Minimum value
230ºF
Maximum value
278ºF
STE Time – Required sterilization time for the process
Entry code
1 or 11
Resolution
1 minute
Minimum value
1 minutes
Maximum value
99 minutes
Dry Time – Required drying time for the process:
Entry code
1 or 11
Resolution
1 minute
Minimum value
Ø minutes
Maximum value
99 minutes
Heat T.O. – Maximum time for heating the chamber for starting sterilization
When the duration of the heating stage is longer than Heat T.O., the program will
collapse with the message ‘Low Heat’.
Entry code
11
Resolution
10 seconds
Minimum value
1200 seconds (20 minutes)
Maximum value
3600 seconds (60 minutes)