Page 48 of 101 pages
Wtr Time – Time for entering the water into the generator after the autoclave
is touching the water –( not applicable)
This value defines in seconds the time that the water pump continues working and
entering water to the steam generator after the upper electrode touches water,
to prevent a situation wherein immediately after the electrode touches water the
pump immediately begins operation.
Entry code
11
Minimum value
Ø second
Maximum value
90 seconds
LimitP add – Adding pressure to the maximum pressure
This value defines addition to the maximum pressure that is computed to the
jacket control.
For example: When the desired temperature is 273ºF, the maximum temperature
maintained in the jacket will be adjusted to a temperature of 137ºC (i.e. 48PSi)
When this pressure is not enough, it can be increased by changing the value of
the LimitP add.
If the parameter value remains Ø, the maximum pressure in the jacket will be
332kPa, however if this value will be 10, the maximum pressure will be 49.6PSi.
Entry code
11
Resolution
0.5 psia
Minimum value
Øpsia
Maximum value
7 psia
Calib Code:
This parameter is used to set the calibration code.
Entry code
107
Pulse Pres1 – The pressure in pulse no.1 in the pre-vacuum stage
This defines the first pulse pressure in the pre-vacuum stage.
If it is defined as 100kPa, the system will not enter steam into the chamber;
however the vacuum pump will immediately operate to create a vacuum.
If the value is defined over 100kPa, the system will open the steam valve to the
chamber until the desired pressure is reached.
Entry code
11
Resolution
1psi
Minimum value
14.7 psi
Maximum value
45 psi
Pulse pres2 – The pressure in pulse 2 and onwards in the pre vacuum stage
After the completion of pulse no.1, the system will enter steam by the defined
value in puls press.
Entry code
11
Resolution
1 psi
Minimum value
14.7 psi
Maximum value
45 psi
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 5596 SP-1
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 12: ...Page 7 of 101 pages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 108: ...AVg Z HiZVb HiZg a oZgh dg i Z A Z HX ZcXZh AVWdgVidgn A cZ...