Page 94 of 101 pages
Type File button
This button enables to select target type file (e.g. in the calibration window below,
the file is: "C:\pcs\dbase\LabTn2.DWN"). Pressing this button will display 'Open
file' dialog box. If the program will find a file with the same name in ADMC
directory, in the Inputs table, the Analog inputs names will be displayed in color.
When the program is connected, rows, the program will display the calibration
tool by double clicking on one of the input table rows. With this tool you can
calibrate the selected analog input.
Description of the Calibration tool
Gain text box – use it for manual inserting of the Gain value.
Offset text box – use it for manual inserting of the Offset value.
Min and Max
– limits for gain and offset values. If the ADMC compatible type file
is not found, Min and Max labels will show "N/A".
Note: When Min and Max are N/A, you must be very careful, because the
gain and offset values are not limited in this case and wrong gain or offset
values may be inserted.
Upload button
– Press this button to upload gain and offset values from the
target platform. All 16 analog inputs gain and offset values will be uploaded.