Page 4 of 101 pages
1.2 Standards
Every autoclave meets applicable requirements of the following standards:
1. A.S.M.E. Code, Section VIII Division 1 for unfired pressure vessels.
2. Complies with Underwriters Laboratories (U.L) requirements.
3. Tested according to Standard EN 285.
Tuttnauer is approved for ISO 9001:2000 (Quality Systems) and ISO 13485
(Quality Systems for Medical Devices).
1.3 Specifications
Description
Model 5596 SP-1
W
20’’ (510 mm.)
H
20’’ (510 mm.)
Chamber Dim.
D
38’’ (970 mm.)
Chamber volume
250 l (9 ft
3
)
W
44” (1120mm).
H
75” (1910 mm).
External dim.
D
56” (1430mm).
Degree of protection
IPX 4
Utilities
Steam Supply
1/2" 75 PSI, 100 lbs/hr Peak, 55lbs/hr Averge
Cold water supply
Maximum hardness value 0.7º2.0mmol/l
Recommended temperature 15°C
1/2", 50-80 PSI
8 GPM Peak,
170 GPM Avg.
Compressed air requirements
Pressure 90-110 PSI, 3.5 SCFM
3/8" thread connection.
Electrical data:
Voltage
1pH, 115V/60 Hz.
Consumption 5A
Current
Utility 10A