Error Opening Type file:
File Name
Description:
This message may be displayed during download process, when
the program is trying to open ADMC compatible type file (*.mdb) in order to get
parameters data.
Reason:
The file may be damaged or used by another application.
Tips:
Try to download another file. If operation succeeds, try to find what
application is using the file.if this file is not used by another application, restart
the PC. If the problem continues, probably the files are damaged. In this case try
to replace the file with an identical file from the ADMC installation CD under
'Support' directory.
Invalid Actual High value!
Invalid Actual Low value!
Invalid Read High value!
Invalid Read Low value!
Description:
These messages are displayed when the user is trying to calculate
Gain and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
Wrong number or non- numeric data has been inserted in one of the
text boxes.
Tips:
Type the right Numeric data in the text box.
The Actual High value is not in range!
The Actual Low value is not in range!
The Read High value is not in range!
The Read Low value is not in range!
Description:
This message is displayed when the user is trying to calculate Gain
and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
A wrong number has been inserted in one of the text boxes. A right
number must be: 1<Number<1000.
Tips:
Type the right number that is included in the above range.
Read High must be equal or great than Read Low!
Description:
This message is displayed when the user is trying to calculate Gain
and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
Read High value is lees than Read Low value
Tips:
Read High value
must be
greater than Read Low value.
Can not calculate. Invalid values found! …
Description:
This message is displayed when the user is trying to calculate Gain
and Offset by pressing the 'Calculate' button in the 'Calibration' window.
Reason:
Read High value - Read Low value = 0
Tips:
Read High value
must be
greater than Read Low value.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 5596 SP-1
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 12: ...Page 7 of 101 pages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 108: ...AVg Z HiZVb HiZg a oZgh dg i Z A Z HX ZcXZh AVWdgVidgn A cZ...