3 Preparation of a communication line
11
3-4 Configuration of RS-232C and GP-IB communication
JOG switch is moved to the CW position several
times, and page 9 (right screen) are displayed. Right
photograph is the state of initial setting.
The blinking cursor is moved on the figure of B.RATE
by CURSOR button. If the central CHG button is
pushed, baud rate can be chosen from 1200, 2400,
4800, 9600, 19200, and 38400. It is only baud rate that
can set up and other values are being fixed to DATA BIT: 8, STOP BIT: 1, NO FLOW
CONTROL, NO PARITY.
A configuration of GP-IB only sets up address. A cursor is moved on the figure of ADDRESS
and address is set up within the limits of 2 to 31.
IF change is completed, the SETUP button will be pushed, and escape from SETUP screen.
3-5 About delimiter character
The format of command is ASCII data, and the delimiter is CR + LF (0x0d + 0x0a).
After receiving CR + LF, command interpretation begins. When sending commands, “CR +
LF” must be added at the end of the command, though the “CR + LF” are omitted in the
command.