8 Timing out signal output
60
<ch>: 0 ~ F(Hexadecimal) for channel.
Output channel fixed mode when enabled, Set up the output channel by the command below.
Confirmation is possible at any time, but the settings have not reflected that it is not the fixed
mode.
Supplement: For channel switching of the timing signal output channel fixed mode
In timing signal output channel fixed mode, switching of the LCD display channel
is possible by CH SEL switch "SETCH" command, even if it is the remote mode.
Command TMGFX?
Confirm the output channel mode of the timing signal.
Reply
DS/ Output from the channel in the display.
EN/ Output channel fixed mode
Example TMGFX? reply: EN
Remarks
Command
TMGCH
<ch><ch><ch><ch>
Set the channel of the timing signal.
Argument <ch>
channel
In order to TP3 from TP0, must be entered in
hexadecimal notation the channel to output the
signal. Please enter "-" in place that does not change
the display. Can not be set at the same time two or
more the same ch.
Reply Nothing
Example TMGCH48CF Set TP0: CH4, TP1: CH8, TP2 CH12, TP3: CH15.
Remarks
It is effective when the timing signal output channel fixed mode.
At remote mode, before setting and after setting, it is effective when both channels are
stopped.
In remote mode, it is effective when both channels (before setting and after setting)
are stopped.
Command
TMGCH?
Confirm the channel of the timing signal.
Reply <ch><ch><ch><ch> order: TP0,TP1,TP2,TP3 (hexadecimal)
Example TMGCH? Reply: 48CF
Remarks