C-6
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
Table C-2
Electrostatic Classifier Firmware Commands List (select)
Command
Title
Description
RMSTATUS
Read Classifier
Status
RMSTATUS returns the classifier status.
Example: RMSTATUS returns a 32 bit integer.
Bits
0
0 = Using Custom efficiency file
1 = not using custom efficiency file
1-3
0 = Neutralizer not present
1 = 3077(A) present
2 = 3088 present and not ionizing
3 = 3088 present and ionizing
4-6
0 = Detector not present
1 = 3772 present
2 = 3775 present
3 = 3776 present
4 = 3787 present
5 = 3788 present
7-9
0 = DMA not present
1 = 3081A present
2 = 3085A present
3 = Custom
10
0 = USB drive not present
1 = USB drive present
11-12
0 = Impactor not present
1 = 0.457 Impactor present
2 = 0.508 Impactor present
3 = 0.710 Impactor present
13
0 = Classifier last calibration <365 days
1 = Classifier last calibration > 365 days
14
N/A
15
0 = Detector last calibration < 365 days
1 = Detector last calibration > 365 days
16-17
0 = Negative HV only
1 = Negative and positive HV
18-20
Not used
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......