Unpacking and Setting Up
2-13
Installing the Model 3088 Neutralizer
W A R N I N G
The Model 3088 Advanced Aerosol Neutralizer contains an x-ray
source, which may be subject to local regulations. During normal
operation, you will not be exposed to x-ray radiation. Do
not
detach the
x-ray source from the neutralizer tube.
C a u t i o n
When installed, the neutralizer completes a flow path inside the
classifier. If you operate the classifier with an external neutralizer or
without any neutralizer, the provided impactors cannot be used, and
aerosol flow rate cannot be measured by the classifier.
To install the Model 3088 Advanced Aerosol Neutralizer, follow these
instructions:
1. If the flag assembly has already been removed, continue with step 3. If
the flag assembly has not been removed, using a Phillips-head
screwdriver, remove the two screws securing the flag assembly and
save for reuse.
2. Lift out the flag assembly by placing your fingers in the depressions on
either side and pulling up on the
Polydisperse Flow
outlet port. Save
the flag assembly for reuse.
3. Remove the black vinyl end caps from the Model 3088 Neutralizer.
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......