2-26
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
6.
If using the side-support bracket accessory, close the bracket and
replace the clevis pin.
7.
Push the end of the long piece of tubing onto the
Sheath Flow +
inlet
on the classifier. Push the end of the shorter length of tubing onto the
Sheath Flow -
inlet.
8.
If you are operating the classifier with a neutralizer installed, you must
connect the Polydisperse flow. Remove the ¼-inch black conductive
tubing from the Accessories Kit. Using the provided tube cutter, or a
clean razor blade, cut a 21-inch length of tubing.