Instrument Operation
5-41
5. To perform a calibration, set the Sheath flow to the first flow calibration
point by pressing the corresponding Setpoint button. The
Raw
and
Classifier
data fields populate, the
Reference
cell for that row is
activated, and an onscreen keypad appears.
Note
: Typically the calibration is performed from the smallest to the
largest flow setpoint; however, the order in which the reference
flow values are recorded is not important as long as the flows
are stable.
Figure 5-25
Electrostatic Classifier User Calibration Screen
–Sheath Flow Setpoint button
6. Monitor the flow rate shown in the
Classifier
cell for 10 seconds or
until the flow rate stabilizes. Record the value from the reference flow
device (to two decimal places) in the
Reference
field using the
onscreen keypad and press
ENTER
.
7. Repeat steps 5 and 6 for each of the flow setpoints, until you have
entered a valid Reference flow value for each setpoint.
Figure 5-26
Electrostatic Classifier User Calibration Screen
–Sheath Flow Setpoint Values
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......