To clean the inner rod and the inside of the outer tube, follow these
instructions:
1. Power off the classifier and unplug the power cord.
2. Disconnect the high-voltage connector from the classifier.
3. Disconnect all the tubing between the classifier and the 1nm-DMA or
Nano DMA.
4. Depress the black lever and remove the 1nm-DMA or Nano DMA from
classifier.
5. Mark the location of the sheath flow fitting on the base plate to facilitate
reassembly.
6. Loosen the two screws in the DMA base plate and remove the plate.
7. Using a 0.050-inch hex wrench, loosen the grounding set screw.
8. Loosen the two screws in the white plastic DMA base and support the
parts so that they do not fall apart.
9. Place the DMA back on its base and separate the sections between
the stainless steel body and the white plastic base (between the
Bypass and Sheath Flow + flow ports).
10. Lift up the stainless steel housing being careful not to make contact
between the outer case and the center rod.
W A R N I N G
Do not
scratch the rod and the inside of the tube. A small scratch, nick,
or burr can disrupt the DMA electric field and adversely affect DMA
performance.
11. Visually inspect the rod. If it is dirty, clean it and the inside of the
1nm-DMA or Nano
DMA’s outer tube:
a. Place your fingers inside the outer tube, or use a soft rubber sheet
to grip the polished surface.
Do not
scratch the inner finish,
b. Pull the outer tube form the inside case.
c. Wash the collector rod and the inside of the outer tube with a soft
cloth soaked in IPA or a mild solvent.
W A R N I N G
Do not
scratch the rod surface or the inner surface of the outer tube.
Do
not
dent the Dacron screen or the cone edge near the top of the rod. If
you damage the mobility analyzer assembly, contact TSI.
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......