Specifications
A-5
Table A-5
Specifications of the Electrostatic Classifier with Nano DMA
Particle type
Solids and nonvolatile liquids
Particle size range
(classifier mode)
Adjustable from 2 to 150 nm
Sizing Accuracy
±1 % at 100 nm for 10:1 sheath/aerosol flow ratio
Maximum input concentration
10
7
particles/cm
3
at 10 nm
Bypass Flow
Port for external pressure or vacuum (provided by user)
Physical measurements -
Nano DMA Model 3085A
Height .......................................... 20.3 cm (8 in)
Outside diameter ......................... 7.9 cm (3.1 in) excluding ports
Weight ......................................... 2.2 kg (4.9 lb)
Polydisperse aerosol inlet ........... ¼-in OD
(standard)
3
8
-in OD (optional for aerosol flow > 4L/min)
Monodisperse aerosol outlet ....... ¼-in OD
Sheath ports and bypass port .....
3
8
-in OD
Complete instrument
Dimensions (LWH) ...................... 40.5 cm
28 cm
40.5 cm
(16 in
11 in
16 in)
Weight ......................................... 17 kg (38 lb)
Order Model 3082-HVPOS when ordering the Classifier Model 3082 to have your Classifier
shipped with dual, switchable DMA-voltage (negative and positive). When ordered later, the
classifier must be shipped to TSI for installation and calibration.
Radiation shielding (for Neutralizers Model 3077 & 3077A) for installation inside Classifier
Model 3082: Order part number: 6005931
Aerosol conditions, operating conditions and storage conditions are reduced for Model 3088,
refer to Model 3088 specifications for details.
Order Model 3085-HIFLOW top cover with 3/8-in inlet (for Nano DMA Model 3085A) for
aerosol flow rates greater than 4 L/min.
Содержание 3082
Страница 2: ......
Страница 32: ...1 12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 76: ...3 6 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 142: ...5 54 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938 This page intentionally left blank ...
Страница 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Страница 176: ......
Страница 189: ...Theory of Operation B 7 Figure B 4 Flow Schematic for the Electrostatic Classifier with Nano or 1nm DMA ...
Страница 222: ......
Страница 224: ......